Skin Cancer Radiotherapy

Small Skin Cancer, Steep Falloff: HDR Brachytherapy vs GentleCure's Deeper-Reaching X-rays

For a small, superficial skin cancer, you want a full dose in the top few millimeters and as little as possible below. Here is how HDR brachytherapy and GentleCure's superficial x-rays compare, and who oversees each.

Illustration of two patients lying face up: a superficial x-ray cone on the nose sends a blue glow deep into the face, while a Leipzig HDR applicator on the nose gives a shallow orange glow
Matthew Culbert, MD

Matthew Culbert, MD

Board Certified Radiation Oncologist

Published · Updated

View physician profile

Short answer: for a small, superficial skin cancer, what you want is a full dose in the top few millimeters of skin and as little as possible below it. HDR brachytherapy with a Leipzig applicator is built for exactly that: it delivers the dose and then falls off steeply, to half the tumor dose by about 1.1 cm and a quarter by about 2.1 cm. GentleCure's superficial x-rays at 70 and 100 kV are more penetrating. They keep carrying meaningful dose toward the structures beneath the tumor: the eye, the bone and sinuses of the face and, on the forehead or scalp, the skull and the brain under it. Just as important is who is in the room. Every HDR treatment requires a radiation oncologist and a medical physicist by federal rule; GentleCure treatments are given by a radiation therapist in a dermatology office.

What a small, superficial skin cancer needs

Most basal and squamous cell skin cancers treated with radiation are small and thin. In GentleCure's own published series, lesions averaged about 1.2 mm deep and 1.2 cm across. A tumor like that lives in the top few millimeters of skin.

So the goal is simple:

  • A full, curative dose to the tumor, usually prescribed about 3 mm deep to cover the base with a margin.
  • A steep drop below it. Everything deeper, such as cartilage, bone, the eye, the sinuses, or the brain beneath the scalp, is healthy tissue that only needs protecting.

Radiation that keeps going past the tumor treats no extra cancer. It only adds dose to tissue that did not need it.

What "percent depth dose" means

Think of sunlight shining into a pool. The water near the surface is bright, and it gets dimmer the deeper you look. Radiation behaves the same way in skin.

Percent depth dose (PDD) is the brightness at each depth compared with a reference point. In this article the reference is the tumor: 100% means the full prescribed dose 3 mm under the skin. If 1 cm down gets 50%, half the tumor dose reaches 1 cm.

HDR with a Leipzig applicator: built for steep falloff

HDR (high-dose-rate) brachytherapy uses a tiny radioactive iridium-192 source on a thin cable. A computer-controlled machine pushes it through a guide tube into an applicator resting on the skin. For skin cancer, a common choice is the Leipzig applicator.

Cutaway drawing of a Leipzig HDR applicator: a tungsten cup with a side guide tube, an iridium source parked 1.6 cm above the skin, a 1 mm plastic cap, and dose falling from 137% at the skin to 25% at 2.1 cm

From the outside in:

  • A tungsten-alloy cup. Tungsten is a heavy metal that blocks radiation in every direction except straight down through the opening. Openings come in 1, 2, and 3 cm sizes to match the tumor.
  • A guide tube that enters from the side (the horizontal type) or the top (the vertical type).
  • The iridium-192 source, about 3.6 mm long, parked in the center of the cup about 1.6 cm above the skin for a few minutes per treatment, then pulled back into the machine.
  • An open cavity with no filter. Nothing evens out the beam, so the center of the field gets a little more dose than the edges.
  • A 1 mm plastic cap that rests on the skin.

The steep falloff comes from that short distance. Radiation spreads out as it travels, and the dose drops with the square of the distance. With the source only 1.6 cm away, every extra centimeter of depth is a big jump in distance, so the dose fades quickly once it has covered the tumor. To give the full dose at 3 mm, the skin surface receives about 137%.

GentleCure: more penetrating x-rays

GentleCure uses the Sensus SRT-100, an x-ray tube about 15 cm from the skin at the top of a metal cone. The operator picks 50, 70, or 100 kV. Higher voltage makes more energetic x-rays that travel deeper before being absorbed.

Because the tube is far away, the x-rays arrive almost parallel and do not spread out much. They fade mainly because tissue absorbs them, and at 70 and 100 kV that absorption is gradual.

In GentleCure's largest published series of 2,917 lesions, the guideline used in 2017 and 2018 called for 50 kV for tumors less than 1.5 mm deep and 70 kV for deeper ones. Across the series, 74% of lesions received 50 kV, 51% received 70 kV, and 3.9% received 100 kV (some received more than one). In other words, about half of these small lesions were treated with a setting that reaches meaningfully deeper than HDR.

Where each beam fades out

Chart of dose versus depth as a share of the tumor dose for GentleCure 50, 70, and 100 kV and HDR with a Leipzig applicator, showing where each falls to half and to a quarter of the tumor dose

Radiation never stops at a sharp line, so physicists describe where it fades. Two useful markers are the depths where the dose drops to half and to a quarter of the tumor dose:

  • HDR, Leipzig applicator: half by about 1.1 cm, a quarter by about 2.1 cm.
  • GentleCure 70 kV: half at about 1.4 cm, a quarter at about 2.7 cm.
  • GentleCure 100 kV: half at about 1.8 cm, a quarter at about 3.5 cm.
  • GentleCure 50 kV: half by about 0.95 cm, a quarter by about 1.9 cm.

At 2 cm, HDR delivers about 27% of the tumor dose, compared with 35% for 70 kV and 45% for 100 kV. At 3 cm, HDR is down to roughly 16%, while 100 kV still delivers about 30%, nearly twice as much.

To be fair to GentleCure: its 50 kV setting fades slightly faster than HDR in soft tissue. It pays for that at the surface, though. To give the full dose at 3 mm, 50 kV would put about 155% on the skin, compared with about 137% for HDR. And as the bone section below shows, every x-ray setting behaves very differently from HDR once bone is involved.

These figures come from physics measurements of individual machines, a 2 cm HDR applicator and a 2.5 cm x-ray cone. Every unit is measured on its own, and applicator size changes the numbers slightly.

Diagram showing HDR rays spreading out from a source inside a Leipzig applicator while superficial x-rays arrive nearly parallel and are absorbed by tissue

On the face, what sits underneath matters

The face is where most of these cancers occur, and the structures under it are close.

Three side views of a patient lying face up with a skin cancer on the bridge of the nose, comparing how deep the dose reaches with an HDR Leipzig applicator, GentleCure 70 kV, and GentleCure 100 kV

On the nose, nasal bone sits only a few millimeters under the skin, with the nasal cavity behind it, the eyes on either side, and the sinuses deeper still. Each panel above gives the same full dose to a tumor 3 mm deep:

  • HDR fades fastest. By the depth of the eye lens, about 1.5 cm, it is down to about 36% of the tumor dose.
  • GentleCure 70 kV still delivers about 46% at that depth.
  • GentleCure 100 kV delivers about 56%. At 3 cm, near the ethmoid sinuses, it still delivers about 30%, compared with about 16% for HDR.

The eye sits beside the nose rather than directly under the beam, so the real lens dose is lower than these center-line numbers. When a tumor is on the eyelid itself, metal eye shields are used to protect the eye.

On the forehead, temple, or scalp, the skull often lies less than 1 cm under the skin, with the brain just beneath it. A beam that still carries roughly half the tumor dose at 1.4 to 1.8 cm deep, as 70 and 100 kV do, is aimed at the skull. Much of that dose is absorbed by the bone itself, and some continues to the brain. HDR's steeper falloff leaves less to reach either.

Bone absorbs superficial x-rays several times harder

Depth-dose charts are measured in water. Bone behaves differently.

Low-energy x-rays are absorbed much more strongly by bone than by soft tissue, because bone contains heavier elements such as calcium. A 2025 computer model of the SRT-100, checked against measurements, found that bone 3 mm under the skin absorbed roughly four to five times the skin-surface dose. With bone 1.5 cm deep, it still absorbed 1.4 to 2.7 times the surface dose, depending on the setting.

Iridium-192 behaves differently. At its energy, bone absorbs about the same dose as the surrounding soft tissue. For a tumor on the nose, forehead, scalp, or shin, that difference applies to every GentleCure setting, including 50 kV.

Then there is oversight

Depth and bone are physics. Who designs, checks, and delivers the treatment is just as important.

HDR brachytherapy is given only in licensed radiation oncology departments. Because it uses a radioactive source, federal Nuclear Regulatory Commission rules require:

  • an authorized physician (a radiation oncologist) and an authorized medical physicist to be physically present when every treatment begins; and
  • a medical physicist and a trained physician to remain physically present for the rest of every treatment.

Professional physics guidelines also call for the radiation oncologist and the physicist to approve each treatment plan before it is delivered.

GentleCure places superficial x-ray equipment and radiation therapists in dermatology offices. Treatments are given by a radiation therapist, with the dermatologist overseeing care. In the largest published series, providers were encouraged to consult a company-affiliated national radiation oncologist through weekly rounds or email for larger or higher-risk lesions. Superficial x-ray machines are regulated by individual states rather than under the federal rules for radioactive sources, and requirements vary.

That difference matters for exactly the questions this article raises: which depth to treat, which energy to use, how to protect the eye, and how to account for bone. Those are radiation oncology judgments. Our comparison of GentleCure and radiation oncology care covers the oversight question and the published evidence in more detail.

What this means for patients

  • For a small, superficial skin cancer, deeper is not better. A technique with a steep falloff, such as HDR with a Leipzig applicator, treats the tumor and spares what lies beneath.
  • Over bone, superficial x-rays at any setting give the bone several times the skin dose. HDR does not.
  • For thicker tumors, a deeper beam may be needed, and for tumors deeper than about 5 mm, neither technique is ideal. The American Brachytherapy Society notes that surface applicators have difficulty achieving adequate dose beyond about 4 to 5 mm. Electrons, interstitial brachytherapy, or surgery may be more appropriate.
  • Surgery, often Mohs, is still the first choice for most operable skin cancers. Radiation is for people who cannot have surgery or would lose function or appearance from it.

Questions to ask

  • How deep is my tumor, and what depth is my dose prescribed to?
  • How much dose will reach the tissue under my tumor, such as bone, the eye, or the brain?
  • If there is bone close by, has its extra absorption been accounted for?
  • Will a radiation oncologist plan my treatment and see me during it?
  • Will a medical physicist check my plan and the machine?

Questions patients often ask

For a small, superficial skin cancer, is deeper radiation better?

No. A superficial tumor only needs a full dose in the top few millimeters. Radiation that keeps going beyond the tumor adds dose to healthy tissue without treating more cancer.

Where does the HDR dose stop?

Radiation never stops at an exact line, but with a Leipzig applicator aimed at a tumor 3 mm deep, the dose falls to half the tumor dose by about 1.1 cm and to a quarter by about 2.1 cm. GentleCure's 100 kV setting does not fall to a quarter until about 3.5 cm.

Can GentleCure's x-rays reach the eye or brain?

They can reach deeper than HDR. Near the eye, 70 and 100 kV beams deliver roughly half the tumor dose at the depth of the eye lens, which is why eye position and shielding matter. On the forehead, temple, or scalp, the skull lies close under the skin; it absorbs several times the skin dose, and some radiation continues to the brain beneath.

Who supervises each treatment?

Federal rules require an authorized radiation oncologist and an authorized medical physicist to be physically present when every HDR treatment starts, and a physicist and trained physician to stay present throughout. In the GentleCure model, a radiation therapist gives superficial x-ray treatments in a dermatology office, with a company-affiliated radiation oncologist available for consultation.

If iridium gamma rays are more energetic, why does HDR fade faster?

Because the source is so close to the skin. Radiation spreads out as it travels, and with a source only about 1.6 cm away, each extra centimeter of depth is a large jump in distance.

Sources and further reading

  1. Commissioning a 50-100 kV X-ray Unit for Skin Cancer Treatment. Journal of Applied Clinical Medical Physics (2015).
  2. Implementation of Superficial Radiation Therapy (SRT) Using SRT-100 Vision for Non-Melanoma Skin Cancer in a Radiation Oncology Clinic. Journal of Applied Clinical Medical Physics (2023).
  3. A Dosimetric Study of Leipzig Applicators. International Journal of Radiation Oncology, Biology, Physics (2005).
  4. Commissioning and Quality Assurance Procedures for the HDR Valencia Skin Applicators. Journal of Contemporary Brachytherapy (2016).
  5. Aspects of Dosimetry and Clinical Practice of Skin Brachytherapy: The American Brachytherapy Society Working Group Report. Brachytherapy (2015).
  6. Monte Carlo Modeling of Radiation Dose from Radiation Therapy with Superficial X-rays. Journal of Applied Clinical Medical Physics (2025).
  7. Optimum Radiation Source for Radiation Therapy of Skin Cancer. Journal of Applied Clinical Medical Physics (2015).
  8. The Treatment of Non-Melanoma Skin Cancer with Image-Guided Superficial Radiation Therapy: An Analysis of 2917 Invasive and In Situ Keratinocytic Carcinoma Lesions. Oncology and Therapy (2021).
  9. 10 CFR 35.615: Safety Precautions for Remote Afterloader Units. U.S. Nuclear Regulatory Commission, Code of Federal Regulations (2026).
  10. AAPM Medical Physics Practice Guideline 13.a: HDR Brachytherapy, Part B. Journal of Applied Clinical Medical Physics (2025).